I am using the BB60C in a remote Python application.
Does anyone know the remote command/API command to set the external reference level offset?
The Spike SW does have an input field for the reference level offset!
I am running on latest SW from the web-page.
Thanks in advance
LPE
AndrewModerator- This reply was modified 2 weeks, 1 day ago by
Andrew.
The API does not have a equivalent command for reference level offset (RLO). The logic for RLO is done entirely in Spike. Fortunately it is simply offsetting the reference level you set through the API. For example, a RLO of 20dB (implying something like a 20dB attenuator in line) would mean setting the reference level in the API 20dB lower.