Forums › General Discussions › What am I doing wrong ?
- This topic has 2 replies, 2 voices, and was last updated 5 years, 7 months ago by
Anonymous.
- AuthorPosts
AnonymousInactiveI am doing a duplexer Sinclair Q2330E measurement.
I short-circuit the measuring cables and push the “Store Thru” button.
This operation is shown in the picture test1.png
I’m connecting a duplexer.
This operation is shown in the picture test2.png
Why is the measurement over 0dB? Marker 3 shows gain ??What am I doing wrong ?
Artur
Justin CrooksModeratorIt looks like some strong VSWR effects. I would throw a 6 or 10 dB attenuator on the SA and TG and repeat the test. If this does not solve the problem, I would try different cables.
AnonymousInactiveThank you for your response. I did as you suggested.
I connected the 6dB attenuator Anritsu to the TG and the 10dB attenuator to the SA and calibrated.Now works correctly!
Artur