Forums › TG Series Discussions › HowTo Correct UNCALIBRATED Label on Tracking Generator Sweep Display
- This topic has 2 replies, 2 voices, and was last updated 8 years, 2 months ago by
tma.
- AuthorPosts
tmaParticipantGreetings,
I recently did a sweep frequency response of an RF network and saved the display. Later I noticed an UNCALIBRATED label in the bottom left hand corner of the display capture file which is attached here.
I recently updated the Spike software and I think it was the first time I used the tracking generator with that particular computer.
I wonder if someone in this group would get me going in the right direction to correct the problem? Do I need to download calibration data manually since I upgraded Spike to a new version?
Thanks in advance for any help!
tma
Attachments:
You must be logged in to view attached files.
Justin CrooksModeratorThe “uncalibrated” indicator means that a valid “store thru” has not been performed, or a setting was changed after it was performed, invalidating it.
If you set the desired center frequency, span, and reference level, select the appropriate scalar network analysis settings, and then store the thru, this indicator should remain off until you change a setting.
Hopefully this helps.
tmaParticipantGreetings Justin,
Many thanks for your prompt reply!
Best Regards,